Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 Instrukcja Użytkownika Strona 49

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 60
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 48
Annexe2TableauderéglageutilisateurSC
App.2-4
IM 12A01A02-12F-E
Paratre
Valeurspar
faut
Valeursapplicables
glage
utilisateur
*Réf.
Advanced
setup
Settings No action No action, Load
factory settings, Save
user settings, Load
user settings
10.5.1
Tag FLXA21-SC (Max. 12 characters) 10.5.2
Passwords Execute: (blank) (Max. 8 characters) 10.5.3
Commissioning: (blank) (Max. 8 characters)
Date/Time Format YYYY/MM/
DD
YYYY/MM/DD,
MM/DD/YYYY,
DD/MM/YYYY
10.5.4
Date
Time
Communication HART None, HART, PH201G 10.5.5
HART Network address 0 0 à 15
PV (Conduct1-
TC1)
(Resist1-TC1)
SV Temperature1 Conduct1-TC1,
Temperature1,
Concent1-TC1, USP
margin1, Conduct2-
TC1, Temperature2,
Conduct2-TC2, USP
margin2, Resist1-
TC1, Resist1-TC2,
Calculated,
Redundant, Empty
TV Empty,
(Concent1-
TC1)
FV Empty,
(Temperature
2)
PH201G Hold contact: Enabled Disabled, Enabled
Fail contact: Fail+Warn Fail+Warn, Fail only,
Disabled
Factory setup 10.5.6
Display
setup
Dual
display
1st row SC1 SC1, SC2, Calculated.
Redundant
10.6.1
2nd row SC2 SC1, SC2, Calculated.
Redundant, Empty
Individual
display
Main
display
SC1 (2) Conductivity (Resistivity) unit Auto Auto, S/cm, mS/cm,
µS/cm, nS/cm
Resistivity:
Auto, mΩ•cm, Ω•cm,
kΩ•cm, MΩ•cm,
GΩ•cm
1st line Conduct1(2)-
TC1,
(Concent1-
TC1,
Resist1-TC1)
Conduct1(2)-TC1,
Temperature1(2),
Concent1(2)-TC1(2),
Resist1-TC1(2)
2nd line Temperature1
(2)
Conduct1(2)-TC1,
Temperature1(2),
Concent1(2)-TC1(2),
Resist1-TC1(2),
Empty
3rd line Empty,
(Concent1-
TC1)
Additional text (blank) (Max. 12 characters)
Przeglądanie stron 48