Annexe2TableauderéglageutilisateurSC
App.2-3
IM 12A01A02-12F-E
Paramètre
Valeurspar
défaut
Valeursapplicables
Réglage
utilisateur
*Réf.
Error
congura-
tion
Conductivity too high (or Concentration) Warn Off, Warn, Fail 10.3
High limit 250.0 mS 0.00 nS à 500.0 S
Conductivity too low (or Concentration) Warn Off, Warn, Fail
Low limit 1.000 µS 0.000 nS à 500.0 S
Temperature too high Warn Off, Warn, Fail
Temperature too low Warn Off, Warn, Fail
Polarization detect Warn Off, Warn, Fail
Calibr.time exceeded Off Off, Warn, Fail
USP Off Off, Warn, Fail
USP safety margin 0.000 % 0.000 à 100.000 %
Congura-
tion error:
1st comp. matrix Fail Off, Warn, Fail
2nd comp. matrix Fail Off, Warn, Fail
Cioncentration table Fail Off, Warn, Fail
Resisitivity too low Warn Off, Warn, Fail
Low limit 4.000 Ω
0.000 mΩ à 1.000 GΩ
Resisitivity too high Warn Off, Warn, Fail
High limit 1.000 MΩ
0.000 mΩ à 1.000 GΩ
Logbook
congura-
tion
Setting
logbook
Sensor: Error on 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2) 10.4
Error off 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2)
Temp.coeff. Off Off, 1-1 (2-1), 1-2 (2-2)
Air adjust limit Off Off, 1-1 (2-1), 1-2 (2-2)
Adj c.c.low limit Off Off, 1-1 (2-1), 1-2 (2-2)
Adj c.c.high limit Off Off, 1-1 (2-1), 1-2 (2-2)
Reference temp. Off Off, 1-1 (2-1), 1-2 (2-2)
Stabilization time Off Off, 1-1 (2-1), 1-2 (2-2)
mA: mA setup Off Off, 1-1 (2-1), 1-2 (2-2)
Hold xed value Off Off, 1-1 (2-1), 1-2 (2-2)
Burn Off Off, 1-1 (2-1), 1-2 (2-2)
Linear output % Off Off, 1-1 (2-1), 1-2 (2-2)
Damping time Off Off, 1-1 (2-1), 1-2 (2-2)
Erase logbook: 1-1 (2-1) 1-1 (2-1), 1-2 (2-2)
Erase? No No, Yes
Warn if logbook “full”: No No, Yes