IM 1E10C1-E
5-6
5. FUNCTION AND DATA SETTINGS
5.3 Setting Meter Factor
The meter factor is engraved on the data plate of the combined flow tube.
Switch Operation Display Description
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INCSET SHIFT INC
SET SHIFT INC
SET SHIFT INC
Change the display into the setting mode through the password
input mode. (Please refer to Chapter 4 “BASIC OPERATION
PROCEDURES.”)
By touching the SHIFT and INC key, select the parameter No.
b08”.
By touching the SET key, change the display into the data
changing mode.
By touching the SHIFT key twice, the cursor moves to the next
digit.
By touching the INC key, change “0” into “1”.
By touching the SHIFT key, the cursor moves to the next digit.
By touching the INC key, change “0” into “2”.
By touching the SET key twice, the setting is completed.
By touching the SET key again, the display goes back to the
number changing mode.
Defalt is set as 1.0000. Below is an example to set it “1.1200”.
(Setting is now completed)
(See 4.2.1 The initial
procedure to change the
display)