Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 Instrukcja Użytkownika Strona 15

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 19
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 14
<3. Commissioning>
13
IM 12A01A02-12E 4th Edition : Sep. 3, 2013-00
3. Commissioning
QUICK SET-UP NOTES:
a. pH measurement module
Under“measurement“aselectionismadeforpH,ORPorpH&ORP.Theselected
measurement must be in accordance with the sensor wiring. When rH measurement is
requestedpH&ORPmustbechosenonthislevel.TherHmustthenbeselectedinthe
commissioning menu.
b. SC measurement module
Under“measurement”aselectionismadebetweenConductivity,Resistivity,Concentration
or Concentration plus Conductivity. On this level only Conductivity or Resistivity can be
selected. Settings for Concentration measurement must be done in the commissioning
menu.
c. DO measurement module
Under“sensortype”aselectionismadeforGalvanicorPolarographic.Theselectedsensor
type must be in accordance with the sensor wiring. Otherwise the analyzer or sensor can be
damaged.
Przeglądanie stron 14
1 2 ... 10 11 12 13 14 15 16 17 18 19

Komentarze do niniejszej Instrukcji

Brak uwag