Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity Instrukcja Użytkownika Strona 57

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 56
49
IM 12D06D05-01E
A
APPENDICES
B. Calculation of temperature coefficient factor
(with two known conductivity values at different temperatures)
Measure the conductivity of the liquid at two temperatures, one below the reference and above
thereferencetemperaturewiththetemperaturecoefficientsetto0,00%/°Candusethefollowing
equationtocalculateatemperaturecoefficient(α).
K
T
K
ref
=
1+α ( T - T
ref
)
K
2
- K
1
α
=
K
1
( T
2
- T
ref
) - K
2
( T
1
- T
ref
)
K
1
K
2
K
ref
=
=
1+α ( T
1
- T
ref
)
1+α ( T
2
- T
ref
)
K
1
(1 + α ( T
2
- T
ref
)) = K
2
(1 + α ( T
1
- T
ref
))
K
1
+ α ( T
2
- T
ref
) - K
2
+ α ( T
1
- T
ref
) = K
2
- K
1
Where T
1
, T
2
:liquidtemperature(°C)
K
1
: conductivity at T
1
(°C)
K
2
: conductivity at T
2
(°C)
Figure 11-1. Conductivity
Calculation example
Calculate the temperature co-efficient of a liquid from the following data.
Conductivity124.5µS/cmataliquidtemperatureof18.0°Canda
conductivity147.6µS/cmataliquidtemperatureof31.0°C.
Substitutingthedataintheaboveformulagivesthefollowingresult.
147.6 - 124.5
x 100= 1.298 %/C
α =
124.5(31.0 - 25) - 147.6(18.0 - 25)
SetthetemperaturecoefficientintheISC450Gconverter.