IM 12J6C3-F-E
Annexe 2-1
Quality
Inspection
Standard
EXA Series
Model DO202
Dissolved Oxygen Transmitter
ANNEXE 3 Quality Inspection Standard (QIS)
A-3-1. Introduction
This inspection procedure applies to the model
DO202 dissolved oxygen transmitter. There is a
serial number, unique to the instrument, which
is stored in non-volatile memory. Each time the
converter is powered up, the serial number is shown
in the display. An example is shown below, for
details see the Users manual:
A-3-2. General Inspection
Final testing begins with a visual inspection of the
unit to ensure that all the relevant parts are present
and correctly fitted.
A-3-3. Safety Test
Terminal 14 and the (-) minus are interconnected.
This interconnection and the external ground
terminal of the housing are connected to a Voltage
generator (600 VAC RMS) for 1 second. The
leakage current should remain below 4 mA.
The interconnection and the external ground
terminal of the housing are connected to a Voltage
generator (100 VDC). The measured impedance

A-3-4. Accuracy Testing
Our automated testing facility checks the accuracy
of the current input of the instrument using a
calibrated current source to simulate the sensor
current.
A-3-5. Accuracy Testing of all supported
temperature elements
Our automated testing facility checks the input
accuracy of the instrument using a calibrated
variable resistor (decade resistor box) to simulate
the resistance of all temperature elements.
ANNEXE 2 Historique de révision du logiciel
Modifications du logiciel du DO202
A-2-1. Modifications apportées dans la version 1.1
Ajout du capteur de température Pb36 (DO30).
A-2-2. Modifications apportées dans la version 1.2
Modification de “Burn low” en association avec le protocole HART.
Affichage du courant de capteur au lieu du courant compensé.
Réparation de mesure instable en étendue ppb lorsque le courant du capteur est négatif.