IM MX100-74E
CAUTION
• Ifalargepullingforceisappliedtotheinput/outputsignalwiresconnectedtothe
MX100andMW100,theterminalorsignalwiremaybreak.Topreventthisfrom
happening,secureallthewiringcablestotheinstallationpanel.
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsoftheuniversalinputmodules.Doingsocandamagethemodules.
• Maximuminputvoltage
 Voltagerangeof1VDCorless,TC,RTD,andDI(contact):±10VDC
 Voltagerangeof2VDCormore,andDI(LEVEL):±120VDC
• Maximumcommonmodevoltage
 Inputtoground: 600VACrms(50/60Hz)
 Betweenchannels: 250VACrms(50/60Hz)(-H04)
• Maximumcommonmodenoisevoltage
 Betweenchannels: 120VACrms(-M10,“b”terminalsarecommonto
allchannels.Theyareconnectedtogetherinternally.)
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsoftheDCV/TC/DIinputmodule.Doingsocandamagethemodule.
• Maximuminputvoltage
 Voltagerangeof1VDCorless,TC,andDI(contact):±10VDC
 Voltagerangeof2VDCormore,andDI(LEVEL):±120VDC
• Maximumcommonmodevoltage
 Inputtoground: 600VACrms(50/60Hz)
• Maximumcommonmodenoisevoltage
 Betweenchannels: 120VACrms
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsoftheFour-wireRTDresistanceinputmodule.Doingsocandamage
themodule.
• Maximuminputvoltage
 Voltagerangeof1VDCorless,RTD,resistance,andDI(contact):±10VDC
 Voltagerangeof2VDCormore,andDI(LEVEL):±120VDC
• Maximumcommonmodevoltage
 Inputtoground: 600VACrms(50/60Hz)
• Maximumcommonmodenoisevoltage
 Betweenchannels,DCvoltage,DI:120VACrms
 Betweenchannels,RTD,resistance:50VACrms
• WiringoftheStrainInputModule(-NDI)
 Whenconnectingabridgehead,inorderthattheemptyweightofthecable
doesnotexceed5kg,ensurethatthecabledoesnothangdownmorethan1.5
m(thedistancetothefloor).Ifthecablehangslongerthan1.5m,securethe
cabletotheinstallationpanelorsomeotherlocation.
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsofthestraininputmodules.Doingsocandamagethemodules.
• Maximuminputvoltage:±10VDC
• Maximumcommonmodevoltage
 Betweenchannels:30VACrms(50/60Hz)
 Inputtoground:250VACrms(-B12,-B35),30VACrms(-NDI)(50/60Hz)
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsofthepulseinputmodule.Doingsocandamagethemodule.
• Maximuminputvoltage:±10VDC
• Maximumcommonmodevoltage
 Inputtoground:250VACrms(50/60Hz)
• Whenusingthepulseinputmodulewithcontactinput,themeasuredsignal
becomeseasilyaffectedbywiringimpedanceathighspeed.Thecableshould
beapproximately25morlesswhenthepulsewidthis0.05ms,or500morless
at0.5ms.
 Thewiringimpedancevariesdependingonthesuchthingsasthecablelength,
type,andwiringconditions.